LEMON Manuals: Even more car manuals for everyone: 1960-2025
Home >> Lexus >> 2007 >> RX 350 AWD >> Repair and Diagnosis >> Engine Performance >> Engine Control System >> SFI System >> DTC P2420 Evaporative Emission System Switching Valve Control Circuit High >> Typical Enabling Conditions
April 5, 2026: LEMON Manuals is launched! Read the announcement.

Typical Enabling Conditions

TYPICAL ENABLING CONDITIONS

The monitor will run whenever these DTCs are not present None
Atmospheric pressure 70 to 110 kPa (525 to 825 mmHg)
Battery voltage 10.5 V or more
Vehicle speed Less than 2.5 mph (4 km/h)
Ignition switch OFF
Time after key-off 5 or 7 or 9.5 hours
EVAP pressure sensor malfunction (P0450, P0452, P0453) Not detected
EVAP canister purge valve Not operated by scan tool
EVAP canister vent valve Not operated by scan tool
EVAP leak detection pump Not operated by scan tool
Both of the following conditions 1 and 2 are set before key-off -
1. Duration that vehicle driven 5 minutes or more
2. EVAP purge operation Performed
ECT 4.4 to 35°C (40 to 95°F)
IAT 4.4 to 35°C (40 to 95°F)

Key-off monitor sequence 1 to 8 

  1. Atmospheric pressure measurement 
    TYPICAL ENABLING CONDITIONS

    Next sequence is run if the following condition is met -
    Atmospheric pressure change Less than 0.3 kPa (2.25 mmHg) in 1 second
  2. First reference pressure measurement 
    TYPICAL ENABLING CONDITIONS

    Next sequence is run if the following conditions are met Condition 1, 2 and 3
    1. EVAP pressure just after reference pressure measurement start -1 kPa (-7.5 mmHg) or less
    2. Reference pressure -4.85 to -1.057 kPa (-36.384 to -7.929 mmHg)
    3. Reference pressure Saturated within 1 minutes
  3. EVAP canister vent valve close stuck check 
    TYPICAL ENABLING CONDITIONS

    Next sequence is run if the following condition is met -
    EVAP pressure change after vent valve is ON 0.3 kPa (2.25 mmHg) or more
  4. Vacuum introduction 
    TYPICAL ENABLING CONDITIONS

    Next sequence is run if the following condition is met -
    EVAP pressure Saturated within 15 minutes
  5. EVAP canister purge valve close stuck check 
    TYPICAL ENABLING CONDITIONS

    Next sequence is run if the following condition is met -
    EVAP pressure change after purge valve is open 0.3 kPa (2.25 mmHg) or more
  6. Second reference pressure measurement 
    TYPICAL ENABLING CONDITIONS

    Next sequence is run if the following conditions are met Condition 1,2,3 and 4
    1. EVAP pressure just after reference pressure measurement -1 kPa (-7.5 mmHg) or less
    2. Reference pressure -4.85 to -1.057 kPa (-36.384 to -7.929 mmHg)
    3. Reference pressure Saturated
    4. Difference between first reference pressure and second reference pressure Less than 0.7 kPa (5.25 mmHg)
  7. Leak check 
    TYPICAL ENABLING CONDITIONS

    Next sequence is run if the following condition is met -
    EVAP pressure when vacuum introduction was complete Lower than second reference pressure
  8. Atmospheric pressure measurement 
    TYPICAL ENABLING CONDITIONS

    EVAP monitor is complete if the following condition is met -
    Atmospheric pressure difference between sequence 1 and 8 Within 0.3 kPa (2.25 mmHg)